I Quotation Summary
A NCL201210 Bridge Coordinate Measuring Machine
1Set
B. RENISHAW UCC Lite Control System 1Set
C. Computer System
-Minimum configuration: Pentium Ⅳ3.0 GHz/1GB RAM/120GB Hard Disk/128MB Graphic Card /17″Color LCD Monitor/ Windows'XP OS
- HP Color DeskJet printer 1Set
D. Calibrate Ball with Support Base 1Set
E. Software System
- CMM Manager â„¢ (CAD version) Software Package (for Windows XP) 1Set
F User and System Manuals 1Set
G. RENISHAW Probe System 1 Set
- PH10T Renishaw Motorized Probe Head
- TP20 Renishaw Touch Prbe
- Renishaw Probe Styli Kit
NC201210 Performance Specifications
Measuring Strokes: 2000 x 1200 x 1000(mm)
Overall Dimension: 3100 x 2000 x 3400 (mm)
Work envelop : 1700 x 1400 x 1000 (mm)
Max part weight : 3000 kg
Resolution : 0.1 (ïm)
Moving speed : 700mm/s
Accuracy according GB/T16857.2-1997 equal ISO 10 360:
Volumetric length measuring error E =2.9 +4L/1000 ï›ïmï
Volumetric Probing ErrorR = 3.9 ïm
Environment Specifications
Room required specifications
-Average room temperature: 18 - 22 °C
-Maximum thermal gradients:1 °C/hour,2 °C/day,1 °C/meter,
-Air supply system
- (Min. air pressure): 0.7 MPa
- (Min. air consumption) : 0.1 m3/min
Electrical requirements
- (Voltage): 220 Vac  10%
- (Frequency) : 50 Hz
- (Current): 10 A
Allowed Maximum Floor Vibration
- Frequency Range: 0 ï¾ 19 Hz19ï¾75 Hz> 75 Hz
- Maximum Amplitude: 0.0011 mm0.0011ï¾0.002 mm0.002 mm