Syringe Size 10mL, 20mL, 30mL, 50mLApplicated SyringeBrand Build-in 10 brands of syringe, 2 reservisions for user-definedVolume Limit 0.1~9999mL; <1000mL 0.1mL increment; ≥ 1000mL 1mL incrementFlow rate range Syringe 10mL: 0.1~400mL/h; Syringe 20ml: 0.1~600mL/hSyringe 30mL: 0.1~900mL/h; Syringe 50ml: 0.1~1300mL/h<100mL/h 0.1mL/h increment or decrement; ≥ 100mL/h 1mL/h increment or decrementBolus rate Syringe10mL: 400mL/h; 20mL: 600mL/h30mL: 900mL/h; 50mL: 1300mL/hKVO rate 0.1mL/h~1mL/h(adjustalbe); 0.1mL/h increment or decrementAccuracry ± 2%(with approved syringe mechanical accuracy less than 1%)Alarm Occlusionsyringe drop offnear emptycompletionlow battery battery power offAC power loss, Motor Abnormal workingNo actionSyringe moved by accidentPressure sensor errorOther function Syringe Automatic IdentificationBolusThree work modes(time, speed, body weight), acumulated volume display, occlusionRelease, mute, over flow rate range control, AC and DC Auto-switch, residual solvent alarm, Previous settings memoryOcclusion sensitivity Adjustable(High, middle, low)0.02MPa~0.06Mpa, 0.06MPa~0.09Mpa, 0.09MPa~0.12MPaKeys 9 keys and 1 jog-shuttleDisplay LCD DOT-matrix displayLED indicator(battery chargeACalarm)luminaire; Display informaiton LED: For battery charge, AC, alarmLCD: Flow rate, acumulated volume, syringe specification, alarm, fault handling solusionPower supply AC: 220V± 22V/50Hz± 1HzDC: 9.6V± 1.6V(over 3 hours continuse work at the flow rate of 5ml/h when fully charged)Power comsumption 15VASafety classify Class type CF; Enclosure protection level: IPX3Enclosure PC+ABS(UL94; V-0)Weight Less than 2.5KgDimension(mm) 314× 167× 140Working conditions temperature: 5C~40C; Relative humility: 20%~90%(no condensation); Atmospheric pressure: 86KPa~106KPaStorage conditions temperature: -20C~55C; Relative humidity: 10%~93%(no condensation); Atmospheric pressure: 70KPa~106KPa