Technical GuidelinesFlow Rate: 1ml/h ~1300ml/h. More than 1000ml/h, 1ml/h step; Less than 1000ml/h, 0.1 ml/h step. Infusing Volume Error: +5%Preset Infusing Volume: 1ml~9999ml. More than 1000ml, 1ml step; Less than 1000ml, 0.1 ml step. 1ml step. Occlusion Sensitivity: Optional choose from higher to lower. Refer to 5.7. KVO: 1-5ml/h (Adjustable, refer to Appendix C)Power: AC: 220V+22V 50Hz+1HzDC: 9.6V+1.6VPower Consumption Less than 20VABattery running hour: More than 3 hours continuously at 30ml/h with fully charged battery. Working Environment: Environment temperature: 10~30 Relative humidity: 30%~75%Atmospheric pressure: 860 hPa ~ 1060hPaDimension: 233*146*269(mm)Weight: About 3kgSafety Classification: Class I with power supply inside, BF type common equipment3. ApplicationBe applicable to common infusion to vein. Prohibition: Prohibited to infuse blood, insulin, medicine for release pain, chemotherapy and putamen anaesthesia.