Medical Infusion Pump (ZNB-XK)

 
 
单价 面议对比
询价 暂无
浏览 754
发货 China(Mainland)付款后3天内
Payment Terms L/C,D/A,D/P,T/T,Western Union,MoneyGram
过期 长期有效
更新 2014-03-15 03:08
 
联系方式
加关注0

BEIJING KELLYMED CO.,LTD

VIP会员第12年
资料未认证
保证金未缴纳
详细说明

Technical GuidelinesFlow Rate: 1ml/h ~1300ml/h. More than 1000ml/h, 1ml/h step; Less than 1000ml/h, 0.1 ml/h step. Infusing Volume Error: +5%Preset Infusing Volume: 1ml~9999ml. More than 1000ml, 1ml step; Less than 1000ml, 0.1 ml step. 1ml step. Occlusion Sensitivity: Optional choose from higher to lower. Refer to 5.7. KVO: 1-5ml/h (Adjustable, refer to Appendix C)Power: AC: 220V+22V 50Hz+1HzDC: 9.6V+1.6VPower Consumption Less than 20VABattery running hour: More than 3 hours continuously at 30ml/h with fully charged battery. Working Environment: Environment temperature: 10~30 Relative humidity: 30%~75%Atmospheric pressure: 860 hPa ~ 1060hPaDimension: 233*146*269(mm)Weight: About 3kgSafety Classification: Class I with power supply inside, BF type common equipment3. ApplicationBe applicable to common infusion to vein. Prohibition: Prohibited to infuse blood, insulin, medicine for release pain, chemotherapy and putamen anaesthesia.

举报收藏 0
更多>本企业其它产品
Medical Feeding Pump Patented Product (ZNB-XY1) Enteral Feeding Pump (ZNB-XY1) Syringe Pump (KL-602) Syringe Pump (KL601) Multi Channel Syringe Pump (KL601) Syringe Pump (KL602)
Home  |  Terms & Conditions  |  About us  |  Contact us  |  Agreement  |  隐私政策  |  Copyright Policy  |  Sitemap  |  Promotion  |  Advertisement  |  Gift  |  Guestbook  |  RSS  |  Report  |  浙ICP备2022009825号-2